VITROS Valproic Acid (VALP) Reagent, REF/Catalog No. 6801710 --- Ortho- Clinical Diagnostics Rochester, NY USA -- Ortho- Clinical Diagnostics High Wycombe, UK --- An in vitro diagnostic. For in vitro diagnostic use only. Used to quantitatively measure valproic acid (VALP) concentration in human serum and plasma.
Recall
- Recall Number
- Z-0984-2012
- Event Number
- 60800
- Firm
- Ortho-Clinical Diagnostics
- FEI Number
- 1000136573
- Product Code
- LEG
- Status
- Terminated
- Root Cause
- Manufacturing material removal
- Initiated
- May 10, 2011
- Posted
- February 8, 2012
- Terminated
- February 12, 2016
- Address
- 100 Indigo Creek Drive, Rochester, NY, 14626
Description
VITROS Valproic Acid (VALP) Reagent, REF/Catalog No. 6801710 --- Ortho- Clinical Diagnostics Rochester, NY USA -- Ortho- Clinical Diagnostics High Wycombe, UK --- An in vitro diagnostic. For in vitro diagnostic use only. Used to quantitatively measure valproic acid (VALP) concentration in human serum and plasma.
Results using VITROS Chemistry Products VALP Reagent were negatively biased compared to other manufacturers. OCD confirmed that patient results were also affected by this issue.
All domestic end-user consignees were initially sent "Important Product Correction Notification" recall communications (dated 5/10/11) by Federal Express overnight mail on 5/10/11. The letters informed users of the issue, and advised them to discontinue using their VITROS VALP reagent until adjusted calibrator values are provided for GEN 14 of VITROS VALP Reagent. The adjusted calibrator values for VITROS VALP Reagent were sent to customers via Assay Data Disk (ADD) Data Release Version (DRV) 5632. Customers were to refer to the Technical Support Communication for details. Users were to confirm receipt of the notice by completing and returning the Fax Reply Form - Confirmation; confirm that all customers or distributors were informed of the issue and instructed to discontinue use of affected product; and lastly confirm that all customers or distributors were sent adjusted calibrator values.
Worldwide Distribution -- US, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.
Approximately 5800 units worldwide (domestic - 4549, foreign - 1287)