FDA Recall Terminated

DePuy DNP Anatomic Fracture Repair System, DNP Anatomic Left, Depuy, Warsaw, IN; REF DNPAL. The device is used in the treatment of fractures and osteotomies of the distal radius.

Recall: Z-0983-2009 · Initiated December 23, 2008

Recall

Recall Number
Z-0983-2009
Event Number
50700
Firm
Depuy Orthopaedics, Inc.
FEI Number
1818910
Product Code
HRS
Status
Terminated
Root Cause
Device Design
Initiated
December 23, 2008
Posted
February 11, 2009
Terminated
October 18, 2010
Address
700 Orthopaedic Dr, Warsaw, IN, 46582-3994

Description

DePuy DNP Anatomic Fracture Repair System, DNP Anatomic Left, Depuy, Warsaw, IN; REF DNPAL. The device is used in the treatment of fractures and osteotomies of the distal radius.

Reason

The DNP nail plate may not attach to the alignment guide (jig) instrument, making it difficult to insert the nail plate into position or to drill needed holes accurately.

Action

The firm's sales force was instructed via "Urgent Device Correction" letter issued December 29, 2008 to inspect all of these instruments in their possession, and those instruments at their hospital customers, and to leave a copy of the letter with each hospital customer who has the instruments on site. The letter describes the issue and corrective actions for customers. Please direct questions about this recall to Depuy Orthopaedics, Inc. at 888-783-7156.

Distribution

Nationwide Distribution.

Quantity

Approx. 3,250 of all products