FDA Recall Terminated

TaqMan HBV Analyte Specific Reagent, Material Number 3322092018, containing HBV Quantitation Standard (HBV QS).

Recall: Z-0983-05 · Initiated February 14, 2005

Recall

Recall Number
Z-0983-05
Event Number
31161
Firm
Roche Molecular Systems Inc
FEI Number
2243471
Product Code
MVU
Status
Terminated
Root Cause
Other
Initiated
February 14, 2005
Posted
July 13, 2005
Terminated
October 3, 2006
Address
1080 Us Highway 202 S, Somerville, NJ, 08876-3733

Description

TaqMan HBV Analyte Specific Reagent, Material Number 3322092018, containing HBV Quantitation Standard (HBV QS).

Reason

Low level of recombinant HIV DNA found in one lot of HBV QS which is a component of one lot of the TaqMan HBV ASR.

Action

Notification letter to LabCorp wa sent on 2/14/2005 via FedEx.

Distribution

One customers received 14 kits on 1/3/2005. This one customer was LabCorp, 1912 Alexander Drive, Research Triangle Park, NC 27709.

Quantity