V*CARE, Infant Resuscitator, with 40 cm HO Pressure Limiting System & Pressure Port Adapter, VN 3000 Series, Non-Sterile. The product included a clear plastic bag with white background which included Direction for Use statement #5. To remove vomitus: Disconnect resuscitator from patient. Tap the patient valve several times while squeezing the bag body. Retest the resuscitator for proper functioning. VentLab Corporation, 2934 Highway 601N, Mockville, NC 27028.
Recall
- Recall Number
- Z-0982-03
- Event Number
- 26369
- Firm
- Ventlab Corporation
- FEI Number
- 1314417
- Product Code
- BTM
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 15, 2003
- Posted
- July 8, 2003
- Terminated
- August 7, 2003
- Address
- 2934 Hwy 601, North Mocksville, NC, 27028
Description
V*CARE, Infant Resuscitator, with 40 cm HO Pressure Limiting System & Pressure Port Adapter, VN 3000 Series, Non-Sterile. The product included a clear plastic bag with white background which included Direction for Use statement #5. To remove vomitus: Disconnect resuscitator from patient. Tap the patient valve several times while squeezing the bag body. Retest the resuscitator for proper functioning. VentLab Corporation, 2934 Highway 601N, Mockville, NC 27028.
Printed bag with instructions to verify the pressure with a 'certified manometer' was omitted in some of the product.
Consignee was notified by fax on 3/15/2003. All product on hand were returned to Ventlab.
The product was distributed to one consignee in CA. The cases were returned to the recalling firm.
9 cases (10 units per case)