FDA Recall Open, Classified

ORAcollect Dx, Catalog Number/SKU: OCD-100, OCD-100A. Material numbers: OCD-100, OCD-100.008, OCD-100.012, OCD-100.014+ACP078, OCD-100.014+ACP078, OCD-100.032 and OCD-100A, OCD-100A.022, OCD-100A.023, OCD-100A.024, OCD-100A.025

Recall: Z-0981-2022 · Initiated February 18, 2022

Recall

Recall Number
Z-0981-2022
Event Number
89831
Firm
DNA Genotek Inc. 3000-500 Palladium Dr Kanata Canada
FEI Number
3003742580
Product Code
OYJ
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
February 18, 2022

Description

ORAcollect Dx, Catalog Number/SKU: OCD-100, OCD-100A. Material numbers: OCD-100, OCD-100.008, OCD-100.012, OCD-100.014+ACP078, OCD-100.014+ACP078, OCD-100.032 and OCD-100A, OCD-100A.022, OCD-100A.023, OCD-100A.024, OCD-100A.025

Reason

Evaporation of stabilizing liquid reducing shelf life.

Action

Notification emailed to customers 2/18/2022. The shelf-life "Collect sample by/Use by" date found on the collection device tube label has been reduced by 12 months. Contact firm if unable to recover the minimum input volume required for sample extraction. Submit online acknowledgement form and contact your DNA Genotek Account manager or Technical Services with questions: [email protected].

Distribution

Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, WI and the countries of Australia Belgium, Brazil, Canada, Chile, China, Croatia Denmark, Finland, Germany, Great Britain, Italy, Japan, Netherlands, South Africa, and Thailand.

Quantity

2,375,514 (US: 2,148,222 OUS: 227,292)