FDA Recall Terminated

Needle Counters; Sterile; Foam Strip; 12.2 x 10.5 x 1.7 cm; 30 Count/60 Capacity; Sterile; Foam Strip; Magnetic Area (Cat. No. 3FS60SSA) Needle Counters; Sterile; Foam Strip; 12.2 x 10.5 x 1.7 cm; 30 Count/60 Capacity; Sterile; Double Foam Strips; (Cat. No. 3FS61SSA) Needle Counters; Sterile; Foam Strip; 12.2 x 10.5 x 1.7 cm; 30 Count/60 Capacity; Sterile; Double Numbered Magnets (Cat. No. 3MM60SSA) Product Usage: Disposal system for counting of used surgical blades and needles.

Recall: Z-0981-2013 · Initiated December 18, 2012

Recall

Recall Number
Z-0981-2013
Event Number
63951
Firm
Cardinal Health
FEI Number
3001236905
Product Code
FSM
Status
Terminated
Root Cause
Package design/selection
Initiated
December 18, 2012
Posted
March 28, 2013
Terminated
August 1, 2016
Address
1430 Waukegan Rd, Mc Gaw Park, IL, 60085-6726

Description

Needle Counters; Sterile; Foam Strip; 12.2 x 10.5 x 1.7 cm; 30 Count/60 Capacity; Sterile; Foam Strip; Magnetic Area (Cat. No. 3FS60SSA) Needle Counters; Sterile; Foam Strip; 12.2 x 10.5 x 1.7 cm; 30 Count/60 Capacity; Sterile; Double Foam Strips; (Cat. No. 3FS61SSA) Needle Counters; Sterile; Foam Strip; 12.2 x 10.5 x 1.7 cm; 30 Count/60 Capacity; Sterile; Double Numbered Magnets (Cat. No. 3MM60SSA) Product Usage: Disposal system for counting of used surgical blades and needles.

Reason

Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.

Action

Cardinal Health sent an Urgent Product Recall letter dated December 18, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immedaitely stop using affected product and discard any product that may be on hand, return the enclosed acknowledgement form, notify any customers to whom you may have distributed product affected by this recall, and contact the appropriate Customer Service group to arrange for credit of any affected product discarded. For questions contact Cardinal Health Professional Services at 800-292-9332.

Distribution

Worldwide Distribution - USA (Nationwide) and the countried of Australia, Canada and Japan.

Quantity

Cat. No. 3FS60SSA: 7,560; Cat. No. 3FS61SSA: 14,184; Cat. No. 3MM60SSA: 12,384