Siemens Syngo based WS 3rd Party V&R medical charged-particle radiation therapy system, Part No: 7345437. Siemens AG, Concord, CA.
Recall
- Recall Number
- Z-0977-2008
- Event Number
- 46200
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2937457
- Product Code
- IYE
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- October 2, 2007
- Posted
- April 30, 2008
- Terminated
- December 24, 2008
- Address
- 4040 Nelson Ave, Concord, CA, 94520-1200
Description
Siemens Syngo based WS 3rd Party V&R medical charged-particle radiation therapy system, Part No: 7345437. Siemens AG, Concord, CA.
Image Orientation Incorrect (Image Shifts) - 1) if the reference image has been calibrated for centering and the reference image is used for patient positioning, the image will shift when the positioning tools are used. 2) This problem occurs only when non-square reference images are used, such as when images are cropped. At this point, the images will refresh, and the center of the reference im
On 10/03/2007, the firm sent Consignees a Customer Safety Advisory Notice informing them of the two potential issues with the display of reference images used for position correction. Consignees will be informed when a solution to this issue has been validated.
Worldwide Distribution-USA including 48 consignees in the US and 39 outside the US.
1 unit