FDA Recall
Terminated
Diamond Mammographic X-Ray System, type MGX-2000, model 5111771. GE Healthcare P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
Recall: Z-0976-2007
·
Initiated November 16, 2005
Recall
- Recall Number
- Z-0976-2007
- Event Number
- 37651
- Firm
- General Electric Med Systems LLC
- Product Code
- IZH
- Status
- Terminated
- Root Cause
- Other
- Initiated
- November 16, 2005
- Posted
- June 28, 2007
- Terminated
- February 24, 2008
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
Diamond Mammographic X-Ray System, type MGX-2000, model 5111771. GE Healthcare P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
Reason
GE Healthcare has become aware of the absence of thread-locking compound associated with the C-arm fasteners of your Diamond mammographic x-ray system that may impact patient safety.
Action
A Product Safety Notification, dated 05/24/07, was sent to Hospital Administrators, Managers of Radiology and Radiologists. This letter describes the safety issue, instructions to continue the use of the product and states that the device will be corrected by GE Healthcare service engineers.
Distribution
Worldwide, including USA, Germany, France, Italy, South Africa, Finland, Norway, Sweden, Venezuela, Benelux, Japan, China, and the UK.
Quantity
106