FDA Recall Terminated

ACCU-CHEK Compact Meter; catalog number 3183572.

Recall: Z-0976-05 · Initiated May 18, 2005

Recall

Recall Number
Z-0976-05
Event Number
31949
Firm
Roche Diagnostics Corp.
FEI Number
1823260
Product Code
NBW
Status
Terminated
Root Cause
Other
Initiated
May 18, 2005
Posted
July 12, 2005
Terminated
December 12, 2005
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

ACCU-CHEK Compact Meter; catalog number 3183572.

Reason

The meter may self-start without a sample having been applied and report erroneous results without an error message if the nose cover is not properly snapped in place and the undosed test strip is bent upwards while the meter is waiting for the sample, or if excess control solution or blood has been applied to the test strip.

Action

A recall letter dated 5/18/05 was sent to health care professionals informing them of the problem and reqeusting them to provide a letter to their patients alerting them to the issue and how to avoid it. A notice of the problem will be placed in all packages of strips in the future.

Distribution

Nationwide.