FDA Recall Open, Classified

Ingenia Elition X (Product Number 781358) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system

Recall: Z-0974-2024 · Initiated January 6, 2022

Recall

Recall Number
Z-0974-2024
Event Number
93737
Firm
Philips North America
FEI Number
3006648320
Product Code
LNH
Status
Open, Classified
Root Cause
Device Design
Initiated
January 6, 2022
Posted
February 2, 2024
Address
222 Jacobs St, Cambridge, MA, 02141-2296

Description

Ingenia Elition X (Product Number 781358) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system

Reason

Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.

Action

Beginning in January 2022, Philips Field Service Engineers (FSE) visited impacted customer sites to replace the magnetic castor wheels of the HA Flextrack Trolley II with non-magnetic castor wheels. All reachable systems were addressed as of August 10, 2023.

Distribution

Worldwide - US Nationwide distribution in the states of FL and NY. The countries of Denmark, Norway, and Sweden.

Quantity

2