FDA Recall
Terminated
VersaRad/Traumex General Radiographic System.
Recall: Z-0974-04
·
Initiated April 23, 2004
Recall
- Recall Number
- Z-0974-04
- Event Number
- 28995
- Firm
- Fischer Imaging Corporation
- FEI Number
- 1418957
- Product Code
- IYB
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 23, 2004
- Posted
- July 20, 2004
- Terminated
- March 13, 2012
- Address
- 12300 Grant St, Denver, CO, 80241-3120
Description
VersaRad/Traumex General Radiographic System.
Reason
If two of the four T-nuts that connect the transverse rail to the ceiling rail fail, the 'U' arm and transverse rails could fall and cause serious injury.
Action
Consignees were notified by letter on 4/23/2004.
Distribution
Nationwide, including VA facilities in MI, MN and VT. No military consignees. Foreign distribution to Australia, Canada, Denmark, Italy, Korea, Netherlands, Norway, Sweden, United Kingdom.
Quantity
138 units