FDA Recall
Open, Classified
Ingenia 1.5T (Product Number 781396) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
Recall: Z-0973-2024
·
Initiated January 6, 2022
Recall
- Recall Number
- Z-0973-2024
- Event Number
- 93737
- Firm
- Philips North America
- FEI Number
- 3006648320
- Product Code
- LNH
- Status
- Open, Classified
- Root Cause
- Device Design
- Initiated
- January 6, 2022
- Posted
- February 2, 2024
- Address
- 222 Jacobs St, Cambridge, MA, 02141-2296
Description
Ingenia 1.5T (Product Number 781396) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
Reason
Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.
Action
Beginning in January 2022, Philips Field Service Engineers (FSE) visited impacted customer sites to replace the magnetic castor wheels of the HA Flextrack Trolley II with non-magnetic castor wheels. All reachable systems were addressed as of August 10, 2023.
Distribution
Worldwide - US Nationwide distribution in the states of FL and NY. The countries of Denmark, Norway, and Sweden.
Quantity
1