FDA Recall Terminated

Malleable C- Retractor, Part Number 397.232, used to provide retraction of the soft tissues during use of trocar instrumentation. Used in the following systems: the Basic Trocar System, Universal Trocar System, Craniofacial Modular Fixation System, Craniofacial (CMF) Distraction System, MatrixORTHOGNATHIC Plating System, and the Matrix MANDIBLE Plating System.

Recall: Z-0973-2015 · Initiated December 15, 2014

Recall

Recall Number
Z-0973-2015
Event Number
70018
Firm
Synthes, Inc.
FEI Number
3005180112
Product Code
GAD
Status
Terminated
Root Cause
Component design/selection
Initiated
December 15, 2014
Posted
January 14, 2015
Terminated
August 18, 2016
Address
1302 Wrights Ln E, West Chester, PA, 19380-3417

Description

Malleable C- Retractor, Part Number 397.232, used to provide retraction of the soft tissues during use of trocar instrumentation. Used in the following systems: the Basic Trocar System, Universal Trocar System, Craniofacial Modular Fixation System, Craniofacial (CMF) Distraction System, MatrixORTHOGNATHIC Plating System, and the Matrix MANDIBLE Plating System.

Reason

Potential failure and/or corrosion of the internal spring of the retractor, which may have been manufactured from an incorrect material. Surgical delay may occur. Spring particles could transfer from the instrument to the patient, and may lead to an adverse tissue reaction and/or infection.

Action

The firm, DePuy Synthes, sent an "URGENT NOTICE: MEDICAL DEVICE RECALL" letter dated December 15, 2014, to its end users and sales consultants. The letter identified the product, problem, and actions to be taken. The End users/sales staff were instructed to immediately remove affected stock from inventory; call DePuy Synthes at 1-800-479-6329 to obtain a Return Authorization (RA) Number; complete and return the attached Verification Section to DePuy Synthes via fax: 866-827-3421 or Scan/email: [email protected], even if you do not have any of the affected product, and return any affected devices to Synthes. If you have any questions, please call 610-719-5450 or contact your DePuy Synthes sales consultant.

Distribution

Worldwide Distribution: US (nationwide) and internationally to Canada.

Quantity

1762