Atlan, Gas Machine for Anesthesia or Analgesia, Catalog numbers 8621500 (Atlan A350) and 8621600 (Atlan A350 XL)

Recall: Z-0971-2024 · Initiated November 17, 2023

Recall

Recall Number
Z-0971-2024
Event Number
93544
Firm
Draegerwerk Ag & Co. Kgaa Dragerwerk Ag & Co. Kgaa Accounts Payable Moislinger Allee
FEI Number
3001104093
Product Code
BSZ
Status
Open, Classified
Root Cause
Use error
Initiated
November 17, 2023
Posted
February 1, 2024
Address
53 55 Lubeck Germany

Description

Atlan, Gas Machine for Anesthesia or Analgesia, Catalog numbers 8621500 (Atlan A350) and 8621600 (Atlan A350 XL)

Reason

Internal backup battery may fail spontaneously while the Atlan is being operated without AC mains supply, resulting in unexpected shutdown of the device while running on batteries.

Action

Draeger US was notified of the recall by their corporate office in Germany on November 17, 2023. All 50 units imported into the US at the time had not been distributed to end-users and were under the direct control of Draeger, Inc. Therefore, the firm did not issue a recall communication to customers. As a corrective action, an Instructions for Use (IFU) supplement has been added to these affected devices. The supplement outlines additional instructions regarding battery operation and inspection. Future shipments of Atlan devices will have the IFU supplement included with the device shipment.

Distribution

No devices were distributed to end-users (customers) in the US. All devices are currently under the direct control of Draeger, Inc.

Quantity

50 imported to US, not distributed to end users