FDA Recall Terminated

Siemens Coherence AG Therapist, Part No: 5863506, medical charged-particle radiation therapy system with software version 2.1. Siemens AG, Concord, CA.

Recall: Z-0971-2008 · Initiated October 2, 2007

Recall

Recall Number
Z-0971-2008
Event Number
46200
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2937457
Product Code
IYE
Status
Terminated
Root Cause
Software design
Initiated
October 2, 2007
Posted
April 30, 2008
Terminated
December 24, 2008
Address
4040 Nelson Ave, Concord, CA, 94520-1200

Description

Siemens Coherence AG Therapist, Part No: 5863506, medical charged-particle radiation therapy system with software version 2.1. Siemens AG, Concord, CA.

Reason

Image Orientation Incorrect (Image Shifts) - 1) if the reference image has been calibrated for centering and the reference image is used for patient positioning, the image will shift when the positioning tools are used. 2) This problem occurs only when non-square reference images are used, such as when images are cropped. At this point, the images will refresh, and the center of the reference im

Action

On 10/03/2007, the firm sent Consignees a Customer Safety Advisory Notice informing them of the two potential issues with the display of reference images used for position correction. Consignees will be informed when a solution to this issue has been validated.

Distribution

Worldwide Distribution-USA including 48 consignees in the US and 39 outside the US.

Quantity

7 units