FDA Recall Terminated

Immunoassay Premium Plus Controls (3x4x5mL and 12x5mL packages)

Recall: Z-0970-2018 · Initiated August 11, 2017

Recall

Recall Number
Z-0970-2018
Event Number
79076
Firm
Randox Laboratories
FEI Number
3003059701
Product Code
JJY
Status
Terminated
Root Cause
Device Design
Initiated
August 11, 2017
Terminated
March 29, 2019
Address
515 Industrial Blvd, Kearneysville, WV, 25430-2778

Description

Immunoassay Premium Plus Controls (3x4x5mL and 12x5mL packages)

Reason

The ACTH in Immunoassay Premium and Premium Plus quality control does not meet the quoted reconstituted stability claim in the product IFU.

Action

On July 24, 2017 an Urgent notice was issued instructing customers to place a copy a notice in all kits remaining in stock. This notifies customers that they should amend the claims for open stability in the instructions for use, to state use immediately after the 30 minute reconstitution period. Questions or concerns can be emailed to [email protected]

Distribution

Distribution US nationwide.

Quantity

199 Kits