Blanketrol II 48C Hyper-Hypothermia System, Model Numbers: 222 and Model Number: 222R. The Blanketrol II 48C System provides a flow of temperature-controlled water to a heat exchanger. The device, in conjunction with a heat exchanger, is used to warm/cool the fluid that enters and exits a heat exchanger. The conductive heat/cool transfer is accomplished through any type of blood oxygenator/heat exchanger unit. The operator of the unit regulates the desired fluid temperature. Frequent observation of the actual blood/fluid temperature and body temperature is required. The Blanketrol II 48 degree C does not contact the patient or the fluids; the water being circulated by the Hemotherm is separated from the fluid by the heat exchanger. To active temperatures above 42 Degrees C, the user must follow explicit instructions in to the owner's manual in order to activate the higher setting. Without following the explicit instructions, the device can only operate within the normal maximum temperature range of: 42 degrees C. The Blanketrol II 48 Degree C System can be used with adult and pediatric patients for extracorporeal circulation of the patient's limb, entire body, or a chemo medication lavage of the abdominal area. The 48C Blanketrol device is used to lower/raise the blood/fluid temperature and/or maintain blood/fluid temperature, as required, through conductive heat transfer. The conductive heat transfer is accomplished through any type of Blood Oxygenator/Heat Exchanger unit. This device is intended to hyperthermic treatment of AIDS and cancer patients and associated research. The device can be set to reach an upper temperature of: 48 degrees C.
Recall
- Recall Number
- Z-0969-2012
- Event Number
- 60721
- Firm
- Cincinnati Sub-Zero Products Inc
- FEI Number
- 1516825
- Product Code
- DWJ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- October 27, 2011
- Posted
- February 7, 2012
- Terminated
- February 26, 2013
- Address
- 12011 Mosteller Rd, Cincinnati, OH, 45241-1528
Description
Blanketrol II 48C Hyper-Hypothermia System, Model Numbers: 222 and Model Number: 222R. The Blanketrol II 48C System provides a flow of temperature-controlled water to a heat exchanger. The device, in conjunction with a heat exchanger, is used to warm/cool the fluid that enters and exits a heat exchanger. The conductive heat/cool transfer is accomplished through any type of blood oxygenator/heat exchanger unit. The operator of the unit regulates the desired fluid temperature. Frequent observation of the actual blood/fluid temperature and body temperature is required. The Blanketrol II 48 degree C does not contact the patient or the fluids; the water being circulated by the Hemotherm is separated from the fluid by the heat exchanger. To active temperatures above 42 Degrees C, the user must follow explicit instructions in to the owner's manual in order to activate the higher setting. Without following the explicit instructions, the device can only operate within the normal maximum temperature range of: 42 degrees C. The Blanketrol II 48 Degree C System can be used with adult and pediatric patients for extracorporeal circulation of the patient's limb, entire body, or a chemo medication lavage of the abdominal area. The 48C Blanketrol device is used to lower/raise the blood/fluid temperature and/or maintain blood/fluid temperature, as required, through conductive heat transfer. The conductive heat transfer is accomplished through any type of Blood Oxygenator/Heat Exchanger unit. This device is intended to hyperthermic treatment of AIDS and cancer patients and associated research. The device can be set to reach an upper temperature of: 48 degrees C.
The suspect hyper-hypothermia device systems sold to physicians for research and/or the treatment of AIDS or Cancer patients, were not appropriately cleared by the FDA prior to distribution.
CSZ sent an "URGENT: FIELD ACTION NOTICE" letter dated January 20, 2011 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to become familiar with the amended manual page and to get approvals from the Institutional Review Board before further use of the device. Additionally, a response form was enclosed with the letter for customers to complete and return via fax to (513)772-9119. Contact CSZ Service Department at 1-800-989-7373 for questions regarding this notice.
Nationwide Distribution-including the District of Columbia and states of CA, GA, MA, MD, MI, NJ, NY, OH, and UT.
16 of these devices were distributed in the U.S., and two (2 ) outside the U.S.