FDA Recall
Terminated
Datex-Ohmeda Aestiva/5 Active Gas Scavenging System, Part Number 1406-8216-000, used with a Model 7900 Ventillator, on a Aestiva/5 Anesthesia Machine, in a Magnetic Reasonance Suite.
Recall: Z-0969-04
·
Initiated April 16, 2004
Recall
- Recall Number
- Z-0969-04
- Event Number
- 28936
- Firm
- Datex-Ohmeda Inc One Ohmeda
- FEI Number
- 2183066
- Product Code
- CBN
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 16, 2004
- Posted
- July 20, 2004
- Terminated
- November 14, 2005
- Address
- Dr. Box 7550, Madison, WI, 53707
Description
Datex-Ohmeda Aestiva/5 Active Gas Scavenging System, Part Number 1406-8216-000, used with a Model 7900 Ventillator, on a Aestiva/5 Anesthesia Machine, in a Magnetic Reasonance Suite.
Reason
Increased circuit pressure as a result of a misassembled Aestiva/5 Active Gas Scavenging System.
Action
Recall letters, dated April 26, 2004, stated that a firm representative would contact the consignee to schedule a visit to verify the proper function of each device unit.
Distribution
The affected units were distributed nationwide in the United States and worldwide.
Quantity
10 units