FDA Recall Terminated

VITROS 4600 Chemistry System (VITROS 5,1 FS System family member), Catalog Number 6802445, IVD --- Ortho Clinical Diagnostics. The VITROS 4600 Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods).

Recall: Z-0968-2015 · Initiated November 19, 2014

Recall

Recall Number
Z-0968-2015
Event Number
69830
Firm
Ortho-Clinical Diagnostics
FEI Number
1000136573
Product Code
JJE
Status
Terminated
Root Cause
Software design
Initiated
November 19, 2014
Posted
January 14, 2015
Terminated
May 21, 2018
Address
100 Indigo Creek Dr, Rochester, NY, 14626-5101

Description

VITROS 4600 Chemistry System (VITROS 5,1 FS System family member), Catalog Number 6802445, IVD --- Ortho Clinical Diagnostics. The VITROS 4600 Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods).

Reason

Software Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibration.

Action

On 11/19/2014, IMPORTANT PRODUCT CORRECTION NOTIFICATION letter (Ref. CL2014-296, dated 11/19/2014) was sent via FedEx overnight courier or ORTHO PLUS e-Communications and/or US Postal Service Priority Mail (for PO Boxes only) to the US consignees (OCD Direct, Drop-Ship and US Federal Government) to inform them of the issue and advise them of the required actions. Immediately install the most recent ADD using the All Assay Data option. If any of your calibrations are expired, the system will post PZB-005 condition codes. Perform calibrations prior to using those reagents. As expected, calibrations that will expire within one hour will also be flagged. Continue to install the most recent ADD at least once each week using the All Assay Data load option so that the system properly flags expired calibrations. --- Foreign affiliates were informed by e-mail on 11/19/2014 of the issue and instructed to notify their consignees of the issue and required actions.

Distribution

Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.

Quantity

339 Total: USA - 102, Foreign - 237