FDA Recall Terminated

VITROS Chemistry Products Mixing Cup Arrays utilized on VITROS 250 & 350 Chemistry Systems Product Usage: Mixing Cup Array is a disposable, intermediate container used in making automatic (auto)-dilutions on the VITROS 250/350 Chemistry Systems. The VITROS 250 Chemistry System is intended for use in the in vitro quantitative, semi quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Microslides. The VITROS 350 Chemistry System is intended for use in the in vitro quantitative, semi quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides.

Recall: Z-0968-2014 · Initiated October 10, 2013

Recall

Recall Number
Z-0968-2014
Event Number
66498
Firm
Ortho-Clinical Diagnostics
FEI Number
1000136573
Product Code
JRB
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
October 10, 2013
Posted
February 10, 2014
Terminated
May 18, 2018
Address
100 Indigo Creek Dr, Rochester, NY, 14626-5101

Description

VITROS Chemistry Products Mixing Cup Arrays utilized on VITROS 250 & 350 Chemistry Systems Product Usage: Mixing Cup Array is a disposable, intermediate container used in making automatic (auto)-dilutions on the VITROS 250/350 Chemistry Systems. The VITROS 250 Chemistry System is intended for use in the in vitro quantitative, semi quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Microslides. The VITROS 350 Chemistry System is intended for use in the in vitro quantitative, semi quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides.

Reason

Ortho Clinical Diagnostics issued a product correction notification for VITROS Chemistry Mixing Cup Arrays utilized on VITROS 250 and VITROS 350 Chemistry Systems. Increase in Software error due to affected Mixing Cup Arrays.

Action

Ortho Clinical Diagnostics sent an Important Product Correction Notification dated October 10, 2013 to all affected customer. The letter identifed the affected product, problem and actions to be taken. Customers were instructed to complete and return the attached Confirmation of Receipt form. For questions contact the Customer Technical Services at 1-800-421-3311.

Distribution

Worldwide Distribution - US Nationwide and the countries of: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, PR, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain

Quantity

6380 units (each unit contains 48 arrays with 10 microcups)