Spacelabs BleaseSirius Anesthesia Workstation, BleaseFocus Anesthesia Workstation, and CAS I/II Absorber contained in Service Kits Part Number 12200902, 050-0659-00 and 050-0901-00 and any workstation installed with the CAS I/II Absorber. Anesthesia workstation for use in the hospital environment and locations not requiring portability.
Recall
- Recall Number
- Z-0968-2013
- Event Number
- 64535
- Firm
- Del Mar Reynolds Medical, Ltd. 1-2 Harforde Ct., Foxholes Business Park Hertford United Kingdom
- FEI Number
- 3002807992
- Product Code
- CBK
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- January 15, 2013
- Posted
- March 20, 2013
- Terminated
- May 14, 2013
Description
Spacelabs BleaseSirius Anesthesia Workstation, BleaseFocus Anesthesia Workstation, and CAS I/II Absorber contained in Service Kits Part Number 12200902, 050-0659-00 and 050-0901-00 and any workstation installed with the CAS I/II Absorber. Anesthesia workstation for use in the hospital environment and locations not requiring portability.
A defect in CAS I/II Absorbers in the Spacelabs Anesthesia Workstations and Service Kits may cause a condition leading to an increase in the carbon dioxide concentration within the inspired gas being delivered to the patient.
Spacelabs sent an URGENT-MEDICAL DEVICE CORRECTION letter, dated 16 January 2013, to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers until a Spacelabs representative replaces the affected product: 1) Immediately cease the use of BleaseSirius and BleaseFocus Anesthesia Workstations that shipped after 31 October 2012. 2) Immediately cease use of Service Kits Part Number 12200902, 050-0659-00 and 050-0901-00 and any workstation containing a CAS I/II Absorber from such Service Kit. 3) Advise your staff of this situation Customers can call Spacelabs Healthcare at 1-800-522-7025 and select 2 for Technical support.
Worldwide Distribution - USA Nationwide including the states of New Mexico, North Carolina and the countries of Argentina, China, Mexico, and Taiwan.
108 total (17 units in the US and 91 units outside the US)