Datex-Ohmeda Aestiva/5 Active Gas Scavenging System, used with a Model 7100 Ventillator, on a Aestiva/5 Anesthesia Machine.

Recall: Z-0968-04 · Initiated April 16, 2004

Recall

Recall Number
Z-0968-04
Event Number
28936
Firm
Datex-Ohmeda Inc One Ohmeda
FEI Number
2183066
Product Code
CBN
Status
Terminated
Root Cause
Other
Initiated
April 16, 2004
Posted
July 20, 2004
Terminated
November 14, 2005
Address
Dr. Box 7550, Madison, WI, 53707

Description

Datex-Ohmeda Aestiva/5 Active Gas Scavenging System, used with a Model 7100 Ventillator, on a Aestiva/5 Anesthesia Machine.

Reason

Increased circuit pressure as a result of a misassembled Aestiva/5 Active Gas Scavenging System.

Action

Recall letters, dated April 26, 2004, stated that a firm representative would contact the consignee to schedule a visit to verify the proper function of each device unit.

Distribution

The affected units were distributed nationwide in the United States and worldwide.

Quantity

73 units