FDA Recall
Terminated
Bulk 16 FR Valved PTFE Sheath/Dilator, Part Number 10614-011
Recall: Z-0968-03
·
Initiated June 6, 2003
Recall
- Recall Number
- Z-0968-03
- Event Number
- 26475
- Firm
- MedAmicus, Inc
- FEI Number
- 2183787
- Product Code
- DYB
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 6, 2003
- Posted
- July 2, 2003
- Terminated
- July 2, 2004
- Address
- 15301 Highway 55 West, Plymouth, MN, 55447
Description
Bulk 16 FR Valved PTFE Sheath/Dilator, Part Number 10614-011
Reason
Stability testing, after accelerated aging of the device, showed that the handle of device did not meet strength specifications.
Action
The consignees were sent recall letters, dated June 6, 2003 or June 20, 2003, that requested the return of the products.
Distribution
The products were distributed to 4 consignees in Georgia, Missouri, New York and North Carolina. Most of the product was shipped to a distributor in New York which probably shipped to other states. There was no direct distribution to foreign or U.S Government accounts.