FDA Recall Terminated

BrainLAB's VectorVision spine is intended for use as an intraoperative image-guided localization system for minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative image data that is processed by a VectorVision workstation. The system is indicated for any medical condition in which the use of stereoscopic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to a CT, x-ray, or MR-based model of the anatomy.

Recall: Z-0967-2014 · Initiated November 15, 2013

Recall

Recall Number
Z-0967-2014
Event Number
67017
Firm
Brainlab AG Kapellenstrasse 12 Feldkirchen Germany
FEI Number
3002619595
Product Code
HAW
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
November 15, 2013
Posted
February 10, 2014
Terminated
March 13, 2017

Description

BrainLAB's VectorVision spine is intended for use as an intraoperative image-guided localization system for minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative image data that is processed by a VectorVision workstation. The system is indicated for any medical condition in which the use of stereoscopic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to a CT, x-ray, or MR-based model of the anatomy.

Reason

Brainlab has determined that pins of thee Instrument Star Unit ML (Calibration with ICM4) with specific serial numbers might have been made of an incorrect material that is not suitable for the necessary disinfection and sterilization of the product. If corrosion occurs and is not detected by the user, and the device is used during surgery, corroded particles of the pin could, directly or indirect

Action

BrainLab send a A FIELD SAFETY NOTICE / PRODUCT NOTIFICATION letter dated November 15, 2013, to all affected customers. The letter requested that customers verify that they are in possession of one or more of the affected devices, remove the affected devices from use, and that Brainlab will provide replacement parts for the customers. Customers with questions can contact Brainlab by telephone at 1-800-597-5911 or via e-mail at [email protected]. For questions regarding this recall call 800-597-5911.

Distribution

Worldwide Distribution - USA including CA, MN, NC, NY, OH, TX and DC. Internationally to Australia, Chile, Germany, Hungary, Italy, Panama, Qatar, Russia, Thailand, United Arab Emirates.

Quantity

25 units