FDA Recall Terminated

Monaco Radiation Treatment Planning (RTP) System

Recall: Z-0965-2019 · Initiated February 8, 2019

Recall

Recall Number
Z-0965-2019
Event Number
82072
Firm
Elekta, Inc.
FEI Number
1037831
Product Code
MUJ
Status
Terminated
Root Cause
Software design
Initiated
February 8, 2019
Terminated
August 19, 2024
Address
400 Perimeter Center Ter NE, Ste 50, Atlanta, GA, 30346-1227

Description

Monaco Radiation Treatment Planning (RTP) System

Reason

If Improve Target Dose was chosen as an optimization model in a previous treatment session, Monaco will automatically use this optimization model again when proceeding with the online plan adaptation of a completion plan when it should not.

Action

Elekta notified customers on about 2/08/2019 via "URGENT IMPORTANT FIELD SAFETY NOTIFICATION." Instructions included to ensure the optimization model is set to "Reproduce Goal Dose" when creating a completion plan. Customers were also requested to post the recall notice in an accessible place for all users to see, advise the appropriate personnel working with the device of the recall, and to complete and return the Acknowledgement Form. The issue will be resolved in Monaco release 5.40.01.

Distribution

Domestic distribution to Wisconsin, Texas. International distribution to Denmark, England, Germany, Netherlands, Sweden.

Quantity

14 devices