EVOTECH Endoscope Cleaner and Reprocessor System 208V, Product Code: 50004 The EVOTECH ECR Endoscope Cleaner and Reprocessor, a washer/disinfector, is indicated for use with high-level disinfectant CIDEX OPA Concentrate Solution and an enzymatic detergent (CIDEZYME XTRA) to achieve cleaning and high level disinfection of heat sensitive (>60C} semi-critical endoscopes.

Recall: Z-0965-2017 · Initiated September 27, 2016

Recall

Recall Number
Z-0965-2017
Event Number
75290
Firm
Advanced Sterilization Products
FEI Number
3003662624
Product Code
FEB
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
September 27, 2016
Posted
January 9, 2017
Terminated
April 17, 2017
Address
33 Technology Dr, Irvine, CA, 92618-2346

Description

EVOTECH Endoscope Cleaner and Reprocessor System 208V, Product Code: 50004 The EVOTECH ECR Endoscope Cleaner and Reprocessor, a washer/disinfector, is indicated for use with high-level disinfectant CIDEX OPA Concentrate Solution and an enzymatic detergent (CIDEZYME XTRA) to achieve cleaning and high level disinfection of heat sensitive (>60C} semi-critical endoscopes.

Reason

Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor System because the hose may come loose and cause a leak.

Action

ASP has sent out Service Representatives to all impacted customers to inspect the hoses for the EVOTECH Endoscope Cleaner and Reprocessor System and will replace the hose if required. The firm issued a Technical Bulletin dated 8/22/2016. Customers were instructed to complete online training prior to performing the actions described in the bulletin and follow instructions. For orders and questions call 888-783-7723 or email: [email protected].

Distribution

US Nationwide Distribution.

Quantity

107 units