FDA Recall Terminated

Alere Cholestech LDX Analyzer, Model Number: 10-004, 14-874, 14-875, in vitro diagnostic.

Recall: Z-0964-2012 · Initiated November 16, 2010

Recall

Recall Number
Z-0964-2012
Event Number
60391
Firm
Alere San Diego
FEI Number
1000125596
Product Code
CHH
Status
Terminated
Root Cause
Device Design
Initiated
November 16, 2010
Posted
February 6, 2012
Terminated
August 22, 2013
Address
9975 Summers Ridge Rd, San Diego, CA, 92121-2997

Description

Alere Cholestech LDX Analyzer, Model Number: 10-004, 14-874, 14-875, in vitro diagnostic.

Reason

Alere has identified that a bias on results for certain analytes tested on the Alere Cholestech LDX System may exist due to humidity variability.

Action

Alere sent an Updated Urgent Recall Notification Letter in March 2012 to all their customers who are in possession of one or more Cholestech LDX Systems. The updated letter provides the customers with important information about the problem identified and the actions to be taken. Customers are instructed to operate their device in a controlled environment that maintains humidity within 40-60% relative humidity if the device has not been upgraded with ROM pack v3.40. If humidity cannot be monitored, the device is no longer CLIA waived. Customers are instructed to complete the customer verification form. A previous notice from Alere sent in November 2010 instructs users to contact Alere to receive the ROM pack upgrade. Customers may email [email protected] to request an electronic form to complete and return via email. Customers can expect their ROM packs approximately 10 to 14 business days after Alere receives their request. If customers have transferred ownership to another party, they can email Alere at [email protected] with the updated contact information. If customers are not currently using their Cholestech LDX System, it is recommended that the ROM Pack upgrade be installed prior to restarting use of any Cholestech LDX analyzer. Customers with questions regarding this communication can call Alere at (877) 441-7440, option 1 or email at [email protected].

Distribution

Worldwide distribution including USA

Quantity

22,000 meters