FDA Recall Terminated

Vitek 2 reagent cards: a) GN ID, REF 21341, b) GP ID, REF 21342, c) ANC ID, REF 21347, d) AST-GNOR, REF 22000, e) AST-GP67, REF 22226, f) AST-P612, REF 22359, g) AST-XN01, REF 410025, h) AST-N222, REF 413083, i) AST-N223, REF 413110, j) AST-N226, REF 413143, k) AST-N248, REF 413397, l) ASR-GN67, REF 413399, m) AST-GN69, REF 413400, n) AST-GN73, REF 413404, o) AST-GN79, REF 413436, p) AST-GN82, REF 413439, q) AST-N261, REF 413753, r) AST-XN06, REF 413944, s) AST-N275, REF 414313, t) AST-GP75, REF 415670, u) AST-N300, REF 416241, v) AST-N308, REF 413913, w) AST-N314, REF 417027, x) AST-N318, REF 417952, y) AST-P643, REF 418671, z) AST-YS08, REF 420739, aa) AST-N348, REF 420856, bb) AST-ST02, REF 420915, cc) AST-ST03, REF 421040, dd) AST-N355, REF 421351, ee) AST-P654, REF 421912, ff) AST-P655, REF 421913, gg) AST-N371, REF 422024, hh) AST-N372, REF 422241, ii) AST-N379, REF 423052,

Recall: Z-0962-2019 · Initiated January 11, 2019

Recall

Recall Number
Z-0962-2019
Event Number
81934
Firm
Organon Teknika Inc
FEI Number
1950204
Product Code
LON
Status
Terminated
Root Cause
Process control
Initiated
January 11, 2019
Terminated
November 6, 2020
Address
100 Rodolphe St, Bldg 1300, Durham, NC, 27712-9402

Description

Vitek 2 reagent cards: a) GN ID, REF 21341, b) GP ID, REF 21342, c) ANC ID, REF 21347, d) AST-GNOR, REF 22000, e) AST-GP67, REF 22226, f) AST-P612, REF 22359, g) AST-XN01, REF 410025, h) AST-N222, REF 413083, i) AST-N223, REF 413110, j) AST-N226, REF 413143, k) AST-N248, REF 413397, l) ASR-GN67, REF 413399, m) AST-GN69, REF 413400, n) AST-GN73, REF 413404, o) AST-GN79, REF 413436, p) AST-GN82, REF 413439, q) AST-N261, REF 413753, r) AST-XN06, REF 413944, s) AST-N275, REF 414313, t) AST-GP75, REF 415670, u) AST-N300, REF 416241, v) AST-N308, REF 413913, w) AST-N314, REF 417027, x) AST-N318, REF 417952, y) AST-P643, REF 418671, z) AST-YS08, REF 420739, aa) AST-N348, REF 420856, bb) AST-ST02, REF 420915, cc) AST-ST03, REF 421040, dd) AST-N355, REF 421351, ee) AST-P654, REF 421912, ff) AST-P655, REF 421913, gg) AST-N371, REF 422024, hh) AST-N372, REF 422241, ii) AST-N379, REF 423052,

Reason

This notice has been initiated due to potential for compromised test card Top Seal Integrity which could: - yield false resistance for antibiotics on the AST panel - cause a false negative ESBL Test - result in a false positive urea (URE) reaction on ID cards

Action

The recall was issued to bioMrieux subsidiaries and distributors via email notification on Friday, January 11th, 2019. Impacted customers were notified via correction notice (customer letter) issued on 01/19/2019 from each associated subsidiary/distributor onto local letterhead. The letter directed the user to perform a "tug test" to determine if the product is defective. Defective units were to be discarded and a credit issued.

Distribution

Worldwide distribution.

Quantity

129,321 cartons