Stryker COCR V40 Femoral Head; xx-Long (+16mm) Neck Length, 26mm Diameter ; Sterile, Forged Vitallium Alloy, Product Number: 6260-5-526; Howmedica Osteonics Corp. Styrker Ireland; Carrigtwohill county Cork, Ireland. Made in Ireland.
Recall
- Recall Number
- Z-0962-2008
- Event Number
- 45902
- Firm
- Stryker Howmedica Osteonics Corp.
- FEI Number
- 3003070421
- Product Code
- JDI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- November 2, 2007
- Posted
- February 7, 2008
- Terminated
- August 30, 2010
- Address
- 325 Corporate Dr, Mahwah, NJ, 07430-2002
Description
Stryker COCR V40 Femoral Head; xx-Long (+16mm) Neck Length, 26mm Diameter ; Sterile, Forged Vitallium Alloy, Product Number: 6260-5-526; Howmedica Osteonics Corp. Styrker Ireland; Carrigtwohill county Cork, Ireland. Made in Ireland.
Fracture of Stem Neck - The firm recently performed laboratory testing indicates that there may be a potential increased risk of stem neck fractures when a V40 +16mm offset femoral head is mated with certain Accolade TMZF femoral stems in patients classified as overweight or obese.
Stryker Orthopaedics sent an Urgent Product Recall notification letters on November 9, 2007, by Federal Express with return receipt. The letter informed users of potential increased risk of stem neck fracture, when a V40 + 16mm offset femoral head is mated with certain Accolade TMZF femoral stems in patients classified as overweight or obese. The firm has asked that inventory be examined, removed and held in quarantine. A Stryker representative will contact the medical facilities to retrieve the product. If a representative has not contacted the facility to date, it is suggested that they contact Stryker at (201) 831-5972.
Nationwide Distribution
5,599 total units