FDA Recall Open, Classified

InPen App, Model/CFN Number: MMT-8061 (Android Users)

Recall: Z-0958-2026 · Initiated November 13, 2025

Recall

Recall Number
Z-0958-2026
Event Number
97904
Firm
Medtronic MiniMed, Inc.
FEI Number
3003166194
Product Code
NDC
Status
Open, Classified
Root Cause
Software Design Change
Initiated
November 13, 2025
Posted
December 18, 2025
Address
18000 Devonshire St, Northridge, CA, 91325-1219

Description

InPen App, Model/CFN Number: MMT-8061 (Android Users)

Reason

When app is uninstalled and reinstalled, insulin pen software issue causes Choose Notification Style Screen to not show during setup so users can't choose to allow notifications to override phone settings when on mute/Do Not Disturb, and previously set up override permission is deleted, so audible and vibratory notifications not received, which may lead to delayed insulin therapy, hyperglycemia.

Action

On 11/13/2025, correction notices were emailed to customers who were asked to update to version 8.0.1 or higher as soon as possible to resolve this issue. 1.Verify your app version: -On the home screen, navigate to the Settings menu (lower right corner). - Select Help and Support. - Note the App Version displayed. 2. If you are running versions 8.0.0, update to the latest version 8.0.1 or higher via the Google Play Store: - Open the Google Play Store. - Tap your profile icon at the top of the screen, then select Manage apps & device. - Tap Updates available or search for the InPen app. - Tap Update next to the InPen app icon. Electronically acknowledge that you have read and understand this notification and have followed the actions listed For questions, contact the firm's 24-Hour Technical Support line at 1-800-646-4633, option #1. The Customer Confirmation Form can also be emailed to the firm at [email protected]

Distribution

US: IL, AZ, MN, NE, KS, TN, CA, FL, TX, WA, MO, NC, NY, WI, IN, ME, CT, RI, MD, PA, OK, KY, MA, DE, UT, ID, VA, MI, OH, LA, NJ, IA, MT, HI, AR, OR, SC, AL, CO, WY, VT, NV, GA, MS, SD, AK, WV, NH, ND, NM. OUS: Austria, Belgium, Chile, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Ireland, Israel, Italy, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, South Africa, Spain, Sweden, Switzerland, United Kingdom

Quantity

8251