FDA Recall Open, Classified

Medtronic MiniMed 640G Insulin Infusion Pump, Ref#s/Model(s) - MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1751, and MMT-1752 *Not for US Distribution

Recall: Z-0958-2020 · Initiated November 21, 2019

Recall

Recall Number
Z-0958-2020
Event Number
84430
Firm
Medtronic Inc.
FEI Number
3003166194
Product Code
OZP
Status
Open, Classified
Root Cause
Device Design
Initiated
November 21, 2019
Posted
February 7, 2020
Address
18000 Devonshire St, Northridge, CA, 91325-1219

Description

Medtronic MiniMed 640G Insulin Infusion Pump, Ref#s/Model(s) - MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1751, and MMT-1752 *Not for US Distribution

Reason

There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that prevents a proper lock. If the reservoir is not properly locked into the pump, it could lead to over or under delivery of insulin, which could then result in hypoglycemia or hyperglycemia.

Action

On November 21, 2019, Medtronic initiated a recall of MiniMed 600 series insulin pumps with a clear retainer ring design. On October 5, 2021, the firm expanded the recall and issued an "Urgent Field Safety Notification" to all patients and health care providers. The updated notification informs consignees that Medtronic will proactively replace all MiniMed 600 series insulin pumps that have a clear retainer ring design with an equivalent pump that has the updated black retainer ring design. Pumps will be replaced at no charge over the coming months even if the retainer ring is not damaged and regardless of the warranty status of the pump. In addition to providing consignees with this updated recall strategy, the firm asked consignees to take the following actions: ACTIONS REQUIRED BY YOU: 1. Complete and return the attached Confirmation Form to acknowledge that you have reviewed and understood this notification. 2. Medtronic is requesting patients visit [www.medtronicdiabetes.com/PumpRing] and complete the form online, or call us at 1-877-585-0166 to indicate their decision to receive a replacement pump at no charge. Replacement pumps will become available in the coming months and patients will be notified when their pump is ready to ship. Please assist patients in making replacement decision considering the safety risk associated with the retainer ring issue. For additional recall details, please view the following link: https://www.medtronicdiabetes.com/customer-support/product-and-service-updates/notice15-letter

Distribution

US:AK,AL,AR,AZ,BM,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY OUS: France, Germany, UK, Italy, Netherlands, Spain, Sweden, Israel, Norway, Austria, Belgium, Denmark, Czech Republic, Finland, Switzerland, Slovakia, Ireland, Poland, S. Africa, Greece, Hungary, Saudi Arabia, United Arab Emira, New Zealand, Australia, Japan, Mexico, Colombia, Brazil, S. East Asia, Korea, Canada,