Cytomics FC 500 MPL Flow Cytometry System with MXP Software, Software Version 2.1, Automated differential cell counter, Part Number: 626554, Beckman Coulter, Fullerton, CA 92834
Recall
- Recall Number
- Z-0958-2008
- Event Number
- 45945
- Firm
- Beckman Coulter Inc
- FEI Number
- 2050012
- Product Code
- GKZ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- June 29, 2007
- Posted
- April 30, 2008
- Terminated
- December 13, 2012
- Address
- 200 S Kraemer Blvd, Brea, CA, 92822-6208
Description
Cytomics FC 500 MPL Flow Cytometry System with MXP Software, Software Version 2.1, Automated differential cell counter, Part Number: 626554, Beckman Coulter, Fullerton, CA 92834
Data generation error: If the cytosettings is not refreshed or restarted during data importation, the output will include old and new data.
Consignees were notified by an Urgent Product Corrective Action letter sent on 6/29/07. The letter provided details on the data error and instruction on how to avoid the error. Users were asked to retain notification as part of the Quality Systems documentation and return a response form acknowledging receipt. For additional information, contact 1-800-526-7694.
Worldwide Distribution- USA and country of Canada.
516 units in the US; 73 units in Canada