FDA Recall Open, Classified

Abbott SE TactiCath Sensor Enabled Contact Force Ablation Catheter, Bi-D, Curve D-F, Model A-TCSE-DF, Sterile.

Recall: Z-0956-2022 · Initiated March 24, 2022

Recall

Recall Number
Z-0956-2022
Event Number
89872
Firm
Abbott
FEI Number
3005334138
Product Code
OAE
Status
Open, Classified
Root Cause
Process control
Initiated
March 24, 2022
Address
5050 Nathan Ln N, Plymouth, MN, 55442-3209

Description

Abbott SE TactiCath Sensor Enabled Contact Force Ablation Catheter, Bi-D, Curve D-F, Model A-TCSE-DF, Sterile.

Reason

Devices were manufactured with a "First Use Date" preprogrammed in the EEPROM which disables the use of the single-use device.

Action

The recalling firm issued letters dated 3/24/2022 on 3/24/22 via hand-delivery. The letters explained the issue, the impact and associated risks, and the steps to reduce risk. The steps included not using any of the remaining inventory of the affected lots listed in Appendix A, complete and return the accompanying Acknowledgment Form, and return all remaining unused affected devices to the recalling firm. The consignee is informed their Abbott representative can assist in returning these devices and obtaining replacements. The notice is to be forwarded to anyone within their organization who may need to be notified and a copy of the notice should be kept along with the completed Acknowledgment Form for their records.

Distribution

Worldwide distribution - US Nationwide - there was government distribution but no military. The countries of Austria, Belgium, Canada, Denmark, France, Germany, Greece, Israel, Italy, Japan, Netherlands, Norway, Palestine, Poland, Portugal, Russia, Slovakia, Spain, Sweden, Taiwan, and United Kingdom.

Quantity

2,134 devices