FDA Recall Terminated

HR90K Platinum Series Headpiece Model Number AB-7300-001 ADVANCED BIONICS LLC 12740 San Fernando Rd. Sylmar, CA 91342 USA. The HiRes90K headpiece is used in conjunction with speech processors manufactured by Advanced Bionics. The Harmony HiResolution Bionic Ear System and Clarion Multi-Strategy Cochlear Implant are cochlear implant systems designed to provide useful hearing to individuals with severe-to-profound sensorineural hearing loss.

Recall: Z-0956-2009 · Initiated July 24, 2008

Recall

Recall Number
Z-0956-2009
Event Number
50204
Firm
Advanced Bionics LLC
FEI Number
3006556115
Product Code
MCM
Status
Terminated
Root Cause
Employee error
Initiated
July 24, 2008
Posted
February 3, 2009
Terminated
March 23, 2012
Address
12740 San Fernando Rd, Sylmar, CA, 91342-3728

Description

HR90K Platinum Series Headpiece Model Number AB-7300-001 ADVANCED BIONICS LLC 12740 San Fernando Rd. Sylmar, CA 91342 USA. The HiRes90K headpiece is used in conjunction with speech processors manufactured by Advanced Bionics. The Harmony HiResolution Bionic Ear System and Clarion Multi-Strategy Cochlear Implant are cochlear implant systems designed to provide useful hearing to individuals with severe-to-profound sensorineural hearing loss.

Reason

Mistuned headpiece - Advanced Bionics recently became aware of a tuning issue with some Platinum Series and S-Series headpieces shipped from May 2007 through April 2008. Tuning refers to the process through which the headpiece's communication frequency is adjusted to match that of the internal device. Potential clinical symptoms include: -No lock -Intermittent lock -Implant locks, but cann

Action

Advanced Bionics is notifying clinicians of this situation so that they can provide appropriate clinical management. The notification letter sent to clinicians in the United States was sent via FedEx so that delivery information could be confirmed. The letter advises that Advanced Bionics recently became aware of a tuning issue with some Platinum Series and S-Series headpieces shipped from May 2007 through April 2008. Auria and Harmony headpieces are not affected by this issue. Tuning refers to the process through which the headpiece's communication frequency is adjusted to match that of the internal device. Since discovery of this issue firm has conducted an extensive technical analysis showing that the use of these headpieces does not pose any safety risk to the patient. While there has been no increase in the number of headpiece-related complaints, firm advises of the potential clinical symptoms: -No lock -Intermittent lock -Implant locks, but cannot measure impedances -Reduction in battery life, primarily for C1 patients with thick flaps. Per the letter, as long as the system maintains lock, there is no degradation in sound quality or stimulation. Further, the impact on battery life for CII and HiRes 90K patients is minimal. Medical professionals are advised that If they have any patients who display any of these clinical symptoms that they attribute to an affected headpiece, to advise the patient to contact Advanced Bionics Customer Service at 877-244-9541 and they will send them a replacement promptly.

Distribution

Nationwide

Quantity

2,049 units