FDA Recall Terminated

MagNA Pure LC Total Nucleic Acid Isolation Kit - Large Volume; Catalog # 03264793001.

Recall: Z-0956-2007 · Initiated May 14, 2007

Recall

Recall Number
Z-0956-2007
Event Number
37948
Firm
Roche Diagnostics Corp.
FEI Number
1823260
Product Code
NPQ
Status
Terminated
Root Cause
Other
Initiated
May 14, 2007
Posted
June 22, 2007
Terminated
December 17, 2009
Address
9115 Hague Rd, Indianapolis, IN, 46250-0416

Description

MagNA Pure LC Total Nucleic Acid Isolation Kit - Large Volume; Catalog # 03264793001.

Reason

False Positive Results: Product labeling states cross contamination is not an issue for high viral load titer samples, but the firm has determined that a potential sample contamination may not be detected at these levels and could result in false positive results being reported.

Action

Consignees were sent an Urgen Medical Device Correction letter dated 5/14/07 which instructs them to define and establish a lower detection limit, to use UNG to prevent carryover whenever possible, and to confirm low positive results through an independent experiment. A revised device correction letter dated 6/15/07 was subsequently sent to each consignee updating the reason for recall and associated risks.

Distribution

Nationwide.

Quantity

4,201