FDA Recall Open, Classified

Brand Name: Rocket Product Name: Rocket 16Fg Seldinger Catheter Procedure Tray Model/Catalog Number: R54549-16-PK Software Version: N/A Product Description: Chest tube manufactured from PVC, with fenestrations and a barium stripe to the indicated French gauge size. Supplied with a procedure pack to allow percutaneous insertion, supplied with connectors to allow attachment to chest drain bottle devices. Component: N/A

Recall: Z-0954-2026 · Initiated November 7, 2025

Recall

Recall Number
Z-0954-2026
Event Number
97994
FEI Number
3002808026
Product Code
GBX
Status
Open, Classified
Root Cause
Labeling design
Initiated
November 7, 2025
Posted
December 18, 2025
Address
Rocket Medical Plc Industrial Estate Sedling Road Wear Industrial Estate Washington Tyne & Wear United Kingdom

Description

Brand Name: Rocket Product Name: Rocket 16Fg Seldinger Catheter Procedure Tray Model/Catalog Number: R54549-16-PK Software Version: N/A Product Description: Chest tube manufactured from PVC, with fenestrations and a barium stripe to the indicated French gauge size. Supplied with a procedure pack to allow percutaneous insertion, supplied with connectors to allow attachment to chest drain bottle devices. Component: N/A

Reason

IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.

Action

On November 12, 2025, Urgent Medical Device Correction Letters were sent to customers. Actions to be taken: 1. Ensure a copy of this letter and the updated IFU are available to all users or potential users of this device. 2. Complete the FSN acknowledgement form and return via the email provided.

Distribution

Domestic: CO, CT, IL, MA, ME, NH, NJ, NM, OH, RI, SC, TX; International: UAE, Austria, Australia, Canada, Germany, Denmark, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, India, Italy, Jersey, Kenya, Sri Lanka, Malta, Netherlands, Norway, New Zealand, Portugal, Saudi Arabia, Singapore, Slovenia, South Africa

Quantity

840 units