FDA Recall Terminated

BioGX SARS-CoV-2 Reagents for BD MAX System Catalog number (REF) 444213. In vitro diagnostic. SARS-CoV-2 Reagents is a real-time RT-PCR test intended for the qualitative detection of nucleic acid from the SARS-CoV-2 in nasopharyngeal, nasal, mid-turbinate, and oropharyngeal swab specimens, nasopharyngeal wash/aspirate or nasal aspirates obtained from individuals suspected of COVID-19 by their healthcare provider.

Recall: Z-0954-2021 · Initiated January 7, 2021

Recall

Recall Number
Z-0954-2021
Event Number
87178
Firm
Becton Dickinson & Co.
FEI Number
1119779
Product Code
QJR
Status
Terminated
Root Cause
Process control
Initiated
January 7, 2021
Terminated
December 3, 2021
Address
7 Loveton Cir, Sparks, MD, 21152-9212

Description

BioGX SARS-CoV-2 Reagents for BD MAX System Catalog number (REF) 444213. In vitro diagnostic. SARS-CoV-2 Reagents is a real-time RT-PCR test intended for the qualitative detection of nucleic acid from the SARS-CoV-2 in nasopharyngeal, nasal, mid-turbinate, and oropharyngeal swab specimens, nasopharyngeal wash/aspirate or nasal aspirates obtained from individuals suspected of COVID-19 by their healthcare provider.

Reason

Some N1 and N2 master mix pouches packaged within the lots of BioGX SARS-CoV-2 Reagents for BD MAX System may be mislabeled. This could result in a kit containing two pouches of N1 master mix or two pouches of N2 master mix, instead of one pouch of each. If laboratory personnel do not notice that both pouches contain tubes with the same color foil seal and use two of the same tubes, there is a low potential for a false negative SARS-CoV-2 result. A false negative SARS-CoV-2 result could lead to an increased risk of disease progression for the patient and additional spread of SARS-CoV-2 if the patient is not quarantined.

Action

On January 12, 2021, BD issued an Urgent Device Correction notice to customers via letters with FedEx tracking. A customer response form is included with the letter that the customer is instructed to complete electronically, email or fax back to BD. Returned response forms will be reconciled with the customer list.

Distribution

US nationwide distribution.

Quantity

938 kits