FDA Recall
Terminated
Product is a surgical nasal/sinus surgical device labeled as a RAD Frontal Finesse Bur, Reference Number 18-83070
Recall: Z-0953-04
·
Initiated January 29, 2004
Recall
- Recall Number
- Z-0953-04
- Event Number
- 28980
- Firm
- Medtronic Xomed, Inc.
- FEI Number
- 1045254
- Product Code
- GFF
- Status
- Terminated
- Root Cause
- Other
- Initiated
- January 29, 2004
- Posted
- July 20, 2004
- Terminated
- May 3, 2012
- Address
- 6743 Southpoint Dr N, Jacksonville, FL, 32216-6218
Description
Product is a surgical nasal/sinus surgical device labeled as a RAD Frontal Finesse Bur, Reference Number 18-83070
Reason
The curve in the shaft of the bur was only 15 degrees not the labeled 40 degrees.
Action
A Recall notification letter was mailed to each hospital account on January 29, 2004 for each type of device they received which was under recall explaining the error and asking the accounts to remove recalled product from inventory, fax them the response sheet and to return the products.
Distribution
Products were distributed to 23 hospitals and surgical centers in NC, NB, IN, UT, TX, NY, WI, MO, TN, ID, OK, FL, MA, WY, NJ, LA.
Quantity
8 boxes 3 burs per box