FDA Recall Terminated

CONTAK RENEWAL 4 AVT CRT-D (model no. M170 and M175) and CONTAK RENEWAL 4 AVT HE CRT-D (model no. M177 and M179). Cardiac Resynchronization Therapy Defibrillator. Sterilized with gaseous ethylene oxide. Guidant Corporation, 4100 Hamiline Avenue North, St. Paul, MN 55112-5798 USA. CONTAK RENEWAL 4 AVT devices provide both atrial and ventricular tachyarrhythmia and cardiac resynchronization therapies.

Recall: Z-0952-05 · Initiated June 24, 2005

Recall

Recall Number
Z-0952-05
Event Number
32496
Firm
Guidant Corporation
FEI Number
2124215
Product Code
NIK
Status
Terminated
Root Cause
Other
Initiated
June 24, 2005
Posted
June 30, 2005
Terminated
December 31, 2006
Address
4100 Hamline Ave N, Saint Paul, MN, 55112-5700

Description

CONTAK RENEWAL 4 AVT CRT-D (model no. M170 and M175) and CONTAK RENEWAL 4 AVT HE CRT-D (model no. M177 and M179). Cardiac Resynchronization Therapy Defibrillator. Sterilized with gaseous ethylene oxide. Guidant Corporation, 4100 Hamiline Avenue North, St. Paul, MN 55112-5798 USA. CONTAK RENEWAL 4 AVT devices provide both atrial and ventricular tachyarrhythmia and cardiac resynchronization therapies.

Reason

Engineering analysis has determined that the magnetic switch in certain devices may stick in the closed position. When the magnetic switch becomes stuck in the closed position treatment of ventricular or atrial tachyarrhythmias is inhibited, while bradycardia pacing is unaffected.

Action

Firm issued a Dr. Doctor letter dated 06/23/05. Letter informs physicians of the issue, clinical implications and recommendations. UPDATE: An Advisory Update, dated 08/01/05 is being issued to physicians to inform them of a new programmer software application upgrade and the programmer screen alerts. Additionally regulatory approval was given for a new switch component that will be incorporated into all Guidant CRT-D products.

Distribution

worldwide. Including United States, Virgin Islands and countries such as Argentina, Australia, Austria, Barbados, Belgium, Bermuda, Canada, Cayman Islands, Chile, Colombia, Cyprus, Czech Republic, Denmark, Eygpt, Finland, France, Germany, Greece, Guadeloupe, Guam, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Labanon, Luxembourg, Marshall Islands, Mexico, Netherlands, New Zealand, Northern Mariana Islands, Norway, Peru, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom and Venezuela.

Quantity

1423