FDA Recall Open, Classified

therascreen KRAS RGQ PCR Kit (24) Model Number(s): REF 870021 (US IVD)

Recall: Z-0951-2022 · Initiated February 25, 2022

Recall

Recall Number
Z-0951-2022
Event Number
89914
Firm
Qiagen Sciences LLC
FEI Number
3003572099
Product Code
OWD
Status
Open, Classified
Root Cause
Device Design
Initiated
February 25, 2022
Address
19300 Germantown Rd, Germantown, MD, 20874-1415

Description

therascreen KRAS RGQ PCR Kit (24) Model Number(s): REF 870021 (US IVD)

Reason

False positive or false negative G12C (12CYS) mutation result for the KRAS Mutation Status resulting from Flags/Warnings may invalidate the result for one or more individual mutation target of the kit.

Action

Qiagen issued Urgent Medical Device Correction therascreen KRAS RGQ PCR Kit (24), REF 870021 letter on 2/25/22. Letter states reason for recall, health risk and action to take: This notice is only relevant for the analysis of samples originating from non-small cell lung cancer (NSCLC) tissue. For continued use of the therascreen KRAS RGQ PCR Kit (24), check the Flags/Warnings in the Sample Result table for automated interpretation of results. Sample results with one or more of the Flags/Warnings listed in this notice should be regarded as invalid with respect to the KRAS status and KRAS Mutation status. Setup a new PCR run to repeat the affected sample(s). To minimize Flags/Warnings for the controls and samples, strict compliance to the IFU guidance should be followed, including but not limited to: Correct reagent mixing is required, and must be ensured at each mixing step during the assay setup. Placement of the Rotor-Gene Q MDx according to Installation Procedures site requirements Forward this information to all individuals and departments within your organization who are using the therascreen KRAS RGQ PCR Kit (24), REF. 870021. If you are not the end user, please forward this notice to the product end user. Review this notice with your laboratory/medical director. Complete the Acknowledgement of Receipt Form attached to this letter by March 10, 2022 and email it to [email protected]. QIAGEN is revising the Instructions for Use, as described in this notice, to reduce any risk resulting from the calling of a false positive or false negative G12C (12CYS) mutation result for NSCLC samples due to the Flags/Warnings listed in this notice and to emphasize strict compliance to the IFU guidance to minimize Flags/Warning. If you have any questions or concerns, pleas.se contact your local QIAGEN Technical Service Department through any of the following: Telephone: 800 362 7737 Email: [email protected]

Distribution

AZ CA CT FL GA IL IN MA MN NC NY OK OR PA TN TX WV

Quantity

435 kits