FDA Recall
Terminated
ESPRIT VENTILATOR, Model V1000
Recall: Z-0951-03
·
Initiated June 8, 2001
Recall
- Recall Number
- Z-0951-03
- Event Number
- 26432
- Firm
- Respironics California Inc
- FEI Number
- 2518422
- Product Code
- CPK
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 8, 2001
- Posted
- July 1, 2003
- Terminated
- May 11, 2004
- Address
- 2271 Cosmos Ct, Carlsbad, CA, 92009
Description
ESPRIT VENTILATOR, Model V1000
Reason
Material hardness on original design check valve may cause premature failure.
Action
Firm plans to replace all of the check valves in the field. Recall was communicated by issuance of a Field Communication 2001-25 dated 8 June 2001.
Distribution
Throughout the US and to Vietnam,Turkey, Thailand, Spain, Singapore, Puerto Rico, Philipines, Peru, Panama, Pakistan, Mexico, Malaysia, Korea, Japan, Italy, Indonesia, Egypt, Columbia, China, Canada, Brazil, Argentina.
Quantity
749