FDA Recall
Terminated
Labeled in part: I Color Complete / Color Lens, blister packing, 14.5mm. (Additional colors include: 1) Illusion style, Chestnut Brown, Violet and others and 2) 2 Tone style: like Misty Blue among others)Eye contact lens Product Usage - Eye contact lens
Recall: Z-0950-2015
·
Initiated September 10, 2014
Recall
- Recall Number
- Z-0950-2015
- Event Number
- 69419
- Firm
- Intertrade Imports Inc.
- FEI Number
- 3007310352
- Product Code
- LPM
- Status
- Terminated
- Root Cause
- Material/Component Contamination
- Initiated
- September 10, 2014
- Posted
- January 12, 2015
- Terminated
- October 7, 2019
- Address
- 4227 Clinton Ave, Jacksonville, FL, 32207-6852
Description
Labeled in part: I Color Complete / Color Lens, blister packing, 14.5mm. (Additional colors include: 1) Illusion style, Chestnut Brown, Violet and others and 2) 2 Tone style: like Misty Blue among others)Eye contact lens Product Usage - Eye contact lens
Reason
A surveillance sample collected by FDA confirmed the presence of Bacillus Cereus and Bacillus Lentus in the contact lenses. I Color Color Complete brand color contact lenses are unapproved medical devices.
Action
The firm verbally contacted each distributor on September 10, 2014 to notify them about the recall and actions that needs to be taken.
Distribution
Worlldwide Distribution - USA including Texas, California, Florida, and St. Lucia, Barbados and St. Marten.
Quantity
750 units