FDA Recall
Terminated
Uroview 2500 X-ray System
Recall: Z-0946-03
·
Initiated May 12, 2003
Recall
- Recall Number
- Z-0946-03
- Event Number
- 26383
- Firm
- GE OEC Medical Systems, Inc
- FEI Number
- 1720753
- Product Code
- JAA
- Status
- Terminated
- Root Cause
- Other
- Initiated
- May 12, 2003
- Posted
- June 21, 2003
- Terminated
- July 7, 2011
- Address
- 384 Wright Brothers Drive, Salt Lake City, UT, 84116
Description
Uroview 2500 X-ray System
Reason
Uncommanded motion of the urological table is possible should the hand control assembly be exposed to liquids.
Action
A Product Safety Alert letter was sent to all consignees on 5/12/2003 recommending that all hand controls be removed from the units.
Distribution
Nationwide. Foreign distribution to Argentina, Australia, Austria, Canada, Germany, Italy, Japan, Mexico, Russia and Switzerland.
Quantity
263 units