Varian brand Varian Clinac Accelerators with Mark, Millennium MLC 120 (Multileaf Collimator) and BrainLAB m3 MCL, Reference/FSCA CP-05591, Model Number: H53, Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA Intended use: The Varian Millennium 120 Collimator (MLC) is an x-ray collimator designed to be mounted on a C or EX Series Varian Clinac radiation therapy linear accelerator and is intended to shape the x-ray field perimeter. Field shape can either be static (fixed) or dynamic. Dynamic field shapes are controlled as a function of either Clinac dose fraction or gantry angle. The intended use is to assist the radiation oncologist in the delivery of radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation.
Recall
- Recall Number
- Z-0945-2012
- Event Number
- 61012
- Firm
- Varian Medical Systems, Inc. Oncology Systems
- FEI Number
- 2916710
- Product Code
- IYE
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- November 10, 2011
- Posted
- February 3, 2012
- Terminated
- February 19, 2015
- Address
- 911 Hansen Way, Palo Alto, CA, 94304-1028
Description
Varian brand Varian Clinac Accelerators with Mark, Millennium MLC 120 (Multileaf Collimator) and BrainLAB m3 MCL, Reference/FSCA CP-05591, Model Number: H53, Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA Intended use: The Varian Millennium 120 Collimator (MLC) is an x-ray collimator designed to be mounted on a C or EX Series Varian Clinac radiation therapy linear accelerator and is intended to shape the x-ray field perimeter. Field shape can either be static (fixed) or dynamic. Dynamic field shapes are controlled as a function of either Clinac dose fraction or gantry angle. The intended use is to assist the radiation oncologist in the delivery of radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation.
Failure to retract or park the Varian MLC leaves prior to use of the demountable BrainLAB m3 could result in the Varian MLC leaves interfering with the treatment field defined by the BrainLAB m3 without an interlock being asserted.
The firm, Varian Medical Systems, Inc., sent an "URGENT MEDICAL DEVICE CORRECTION-URGENT FIELD SAFETY NOTICE" letter dated November 10, 2011 to all affected users. The letter describes the product, description of problem, details, recommended actions and Varian's actions. Varian Service will contact customers to arrange installation, once modifications are available. ***12-12-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.*** The letter also included special instructions for customers outside the USA and Canada. The customers were instructed to complete and return the attached Proof of Notification or Receipt Verification Card to Varian Medical Systems once they have read the document. If you require further clarification, feel fee to contact your local Varian Customer Support District or Regional Manager at: USA and Canada: 1-888-VARIANS (1-888-827-4265); Europe: +41 41 749 8844; or Email: North America: [email protected]; Australia/New Zealand: [email protected]; Europe: [email protected]; South East Asia: [email protected]; China/Asia: [email protected]; Japan: [email protected]; Latin America: [email protected] or Internet: Oncology Systems customer site-www.myvarian.com and Varian Medical Systems public site-www.varian.com.
Worldwide distribution: USA (nationwide) including Puerto Rico and countries of: Algeria, Austria, Australia, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Malaysia, Mexico, Netherlands, Norway, Philippines, Poland, Portugal, Qatar, Saudi Arabia, Slovenia, South Korea, Spain, Sri Lanka, Sweden, Taiwan, Turkey, United Kingdom and Venezuela.
205 units***12-12-12 AMENDED TO: 213 units***