FDA Recall Terminated

Symbiq One-Channel list number 16026-04 and Two-Channel list number 16027-04 Infusion System, a general purpose infuser, Distributed and Manufactured by Hospira Inc, Morgan Hill, CA. Is a general purpose infuser designed to deliver fluids, solutions, medications, agents, nutritionals, electrolytes, blood and blood products for parenteral, enteral, intravenous, intra-arterial, subcutaneous, epidural, or irrigation routes of administration. Intended primarily for use in the hospital setting and can be used in other acute and non-acute care areas, such as but not limited to Home Care, Nursing Homes, Mobile Intensive Care, Ambulatory Infusion Centers, Hospice, Subacute facilities, Outpatient/Surgical Centers, Long Term Care, Urgent Care, Transport and Physician Offices.

Recall: Z-0944-2009 · Initiated December 3, 2008

Recall

Recall Number
Z-0944-2009
Event Number
50541
Firm
Hospira Inc
FEI Number
2921482
Product Code
FRN
Status
Terminated
Root Cause
Software design
Initiated
December 3, 2008
Posted
February 27, 2009
Terminated
April 3, 2012
Address
755 Jarvis Dr, Morgan Hill, CA, 95037-2810

Description

Symbiq One-Channel list number 16026-04 and Two-Channel list number 16027-04 Infusion System, a general purpose infuser, Distributed and Manufactured by Hospira Inc, Morgan Hill, CA. Is a general purpose infuser designed to deliver fluids, solutions, medications, agents, nutritionals, electrolytes, blood and blood products for parenteral, enteral, intravenous, intra-arterial, subcutaneous, epidural, or irrigation routes of administration. Intended primarily for use in the hospital setting and can be used in other acute and non-acute care areas, such as but not limited to Home Care, Nursing Homes, Mobile Intensive Care, Ambulatory Infusion Centers, Hospice, Subacute facilities, Outpatient/Surgical Centers, Long Term Care, Urgent Care, Transport and Physician Offices.

Reason

Hypersensitivity to normal motor friction/drag may cause malfunction errors and alarms that are designed to stop pump operations.

Action

Recall initiated 12/03/2008. Consignees were notified of the issue via an Urgent: Device Field Correction letter, dated 12/3/2008, and were requested to respond appropriately to any alarm or warnings displayed by the device. The firm is making arrangements to screen each Symbiq pump in customer facilities and implement the appropriate enhancements.

Distribution

Worldwide Distribution --- including USA and countries of Canada and Australia.

Quantity

combined 11,840 units