FDA Recall
Terminated
Supergrade Earloop Masks, surgical mask; 50 masks per box; Distributed by Safco Dental Supply Co., Northbrook, IL 60062
Recall: Z-0944-03
·
Initiated June 3, 2003
Recall
- Recall Number
- Z-0944-03
- Event Number
- 26382
- Firm
- Safco Dental Supply Co. Inc.
- FEI Number
- 1416277
- Product Code
- FXX
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 3, 2003
- Posted
- June 21, 2003
- Terminated
- March 24, 2004
- Address
- 2121 Shermer, Northbrook, IL, 60062
Description
Supergrade Earloop Masks, surgical mask; 50 masks per box; Distributed by Safco Dental Supply Co., Northbrook, IL 60062
Reason
The product labeling incorrectly lists the particle filtration efficiency as exceeding 99% at 0.1 micron when it should state 95% filtration.
Action
Recalled by letter dated 5/30/03, informing the dentists that the earloop masks were mislabled with the wrong particle filtration efficiency. Correction labels were included with the letter for the dentists to field correct the boxes of masks in their office inventory.
Distribution
Nationwide
Quantity
60,000 boxes