FDA Recall Terminated

Optilite Freelite Mx Kappa Free Kit REF LK016.M.OPT.A

Recall: Z-0943-2021 · Initiated November 18, 2020

Recall

Recall Number
Z-0943-2021
Event Number
86926
Firm
The Binding Site Group, Ltd.
FEI Number
3002808340
Product Code
DFH
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
November 18, 2020
Terminated
December 23, 2024
Address
8 Calthorpe Road, Birmingham United Kingdom

Description

Optilite Freelite Mx Kappa Free Kit REF LK016.M.OPT.A

Reason

Due to customer complaints of positive bias impacting quality control. The high and low level controls are demonstrating a positive bias compared to assigned values.

Action

On 11/30/2020, the firm emailed an "Urgent Field Safety Notice" to its U.S. Customers informing them that investigations have confirmed that the quality control (QC), high and low controls, provided with specific product and lot numbers are consistently demonstrating a ~15% positive bias compared to assigned values and customers may experience QCs consistently reporting results close to the upper acceptable limits, and QC failures may occur on the affected products. Customer are instructed: 1. Users may continue to use any remaining kits of the affected product in their stock. As per the product package insert sample results obtained should only be accepted if the control results are within +/-20% of the concentrations stated on the QC certificate provided. 2. To complete the E-Back Form TSWS18 to [email protected] or to their local representative within one week of receiving this notification. Customers are also instructed: -To pass on the notification to all those who need to be aware within their organization or to any organization where the potentially affected devices have been transferred. -To transfer this notification to other organizations on which the information has an impact. -To maintain awareness on this notification for any appropriate period. Any questions, contact the local representative or Technical Support Group: UK technical support is: Telephone: +44(0) 1214569696 Email: [email protected] Customers in the USA may choose to instead contact the local representative at: Telephone: 800-633-4484 [email protected]

Distribution

U.S.: CO, FL, IA, MA, NC, NJ, NY, OH, SD, and UT. O.U.S.: Not provided

Quantity

213 kits in the U.S.