FDA Recall Terminated

Stryker InTouch 1.0 Zoom Critical Care Bed (Model 2140) with the Integrated PositionPro mattress (Model 2920) option. The device is intended to be used as an Intensive Care Unit bed.

Recall: Z-0943-2009 · Initiated January 5, 2009

Recall

Recall Number
Z-0943-2009
Event Number
49610
Firm
Stryker Medical Div. of Stryker Corporation
FEI Number
1831750
Product Code
FNL
Status
Terminated
Root Cause
Device Design
Initiated
January 5, 2009
Posted
February 25, 2009
Terminated
November 5, 2010
Address
3800 E. Centre Ave., Portage, MI, 49002

Description

Stryker InTouch 1.0 Zoom Critical Care Bed (Model 2140) with the Integrated PositionPro mattress (Model 2920) option. The device is intended to be used as an Intensive Care Unit bed.

Reason

The touch screen on the footboard may display a "tilt over range" error, resulting in a precautionary stop of the bed motions which result in the bed appearing to have locked out some of its bed positioning features. Also, the "hold" button, used to interrupt an intended patient turn on beds with an integrated PositionPro mattress, may only function intermittently, resulting in an inability to ho

Action

Consignees were notified of the problems via letter dated January 5, 2009, advised to check their beds and to take them out of service if the foot board touchscreen displayed a "tilt over range" error, and that a Stryker Medical field service representative would contact them to perform software upgrades on the affected beds. Please direct questions to Stryker Medical at 1-269-324-6644.

Distribution

Worldwide Distribution including USA and countries of Canada, Japan and Switzerland.

Quantity

1,288 of all models