ARIA Radiation Therapy Management (RTM) VERSIONS 13.6, 15.1, 15.5, 15.6, 16.0, 16.1
Recall
- Recall Number
- Z-0942-2022
- Event Number
- 89922
- Firm
- Varian Medical Systems Imaging Laboratory GmbH Tafernstrasse 7 DATTWIL Baden Switzerland
- FEI Number
- 3003094912
- Product Code
- IYE
- Status
- Completed
- Root Cause
- Software Design Change
- Initiated
- March 21, 2022
Description
ARIA Radiation Therapy Management (RTM) VERSIONS 13.6, 15.1, 15.5, 15.6, 16.0, 16.1
Software issue for treatment plan and image management application may result in mismatch values which could result in treatment in the wrong location
On 3/21/22, Varian Medical Systems issued a "Urgent Medical Device Correction" notification to affected consignees. In addition to informing consignees about the recall, Varian asked consignees to take the following actions: Please retain a copy of this document along with your most current product labeling. Advise the appropriate personnel working in your radiotherapy department of the content of this letter. This document will be posted to the Varian Medical Systems customer support website: http://www.MyVarian.com. Please complete the attached Return Response form provided. Kindly return the completed form to [email protected].
Worldwide distribution - US Nationwide distribution in the states of MI, KS, OH, VA, DC, FL, PA, OK, MN, AZ and the countries of South Korea, Netherlands, Poland, China.
14