FDA Recall Terminated

INFINITY DUAL HEMO MCable Pod

Recall: Z-0942-2018 · Initiated October 4, 2017

Recall

Recall Number
Z-0942-2018
Event Number
79209
Firm
Draegar Medical Systems, Inc.
FEI Number
3005783425
Product Code
MHX
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
October 4, 2017
Terminated
June 15, 2018
Address
6 Tech Dr, Andover, MA, 01810-2434

Description

INFINITY DUAL HEMO MCable Pod

Reason

Dual Hemo MCable Pods with Revision Index (RI) 15 and 16, which are used with the Drger Infinity Acute Care System, may permit liquid ingress inconsistent with their IPX4 rating, which could potentially result in an incorrect measured value of invasive blood pressure.

Action

In September of 2017 an Urgent medical Device Recall letter was issued to customers informing them of the recall and to request that inventory be checked to see if the product subject in the recall is on the premises. To do this, the customer will check the label for the revision index number which is located on the upper middle portion of label. RI number 15 and 16 are included in the recall. No other numbers are subject to recall. Affected product will receive a sleeve from the company to correct the issue. Questions or concerns can be directed to 1-800-437-2437 (Press 1, then press 32349) or contact Customer Care at 1-800-437-2437 (Press 4, then press 4 again)

Distribution

Nationally

Quantity

2,760